首页 > 器件类别 > 热门应用 > 测试测量

DM-III MULTITEST

power analyzers pwr qual record/anal data logger

器件类别:热门应用    测试测量   

厂商名称:Amprobe

器件标准:

下载文档
器件参数
参数名称
属性值
Manufacture
AMPROBE
产品种类
Product Category
Power Analyzers
RoHS
Yes
类型
Type
Handheld
电流额定值
Current Rating
1 kA
电压额定值
Voltage Rating
50 V to 1 kV
频率
Frequency
50 Hz/60 Hz
Accuracy
+/- 5 %
True RMS
True RMS
Unit Weigh
1.2 kg
文档预览
Data Sheet
DM-III MULTITEST F
Power Quality Recorder
Amprobe’s full-featured Three-Phase Power Quality Recorders
provide the essential functions and capabilities required to
operate accurately and effectively in today’s demanding
electrical environments.
POWER QUALITY ANALYZER/DATA LOGGER
− True RMS (TRMS)
− Measures & Records Broad Spectrum of Power Quality
Parameters
• AC Current
• AC Voltage to 600 V
• Sags and Surges
• Harmonics
• Active, Reactive and Apparent
• Power
No hassle
warranty
No waiting.
No shipping
charges.
Our commitment to
high-quality products
and customer service
is demonstrated by
our industry exclusive
“No Hassle” warranty.
In the unlikely event
that an Amprobe Test
Tool requires warranty
service, any of our
local dealers are
authorized to replace
it, on the spot.
(note: $500 MSLP limit)
• Peak Demand
• Power Factor
• Frequency
• Phase Sequence
− Compatible with wide range of current transducers
− Works with single and three phase
− Detects & records Sags and Surges
− Displacement power factor for power factor correction
determination
− Built in scope displays waveforms
− Phase sequence indication
− Records up to 64 parameters
− Selectable fundamental frequency
− Special data compression system
− Download capabilities, Windows compatible
− A complete kit: 3000A Flexible CT, 1000A Clamp, Voltage
Leads, Ground Probes & Leads, PC software & cable
continued on next page
600V
Amprobe® Test Tools
www.Amprobe.com
Data Sheet
INSULATION TESTER FUNCTIONS:
− Tests insulation integrity of wires, cables, transformers &
electrical motors
− Selectable test voltages up to 1000 V
− Programmable timer to perform the Dielectric Absorption Ratio
Test
− Sensitive Ohmmeter for checking resistance of motor windings
− Selectable polarization of ohmmeter for checking grounding
continuity
− Automatic voltmeter protects against misuse on hazardous
energized systems
GROUND RESISTANCE & RESISTIVITY FUNCTIONS:
− Three measuring modes:
• 2 point continuity/resistance test
• 3 point Fall of Potential test
• 4 point Earth Resistivity measurement
− Automatic voltage measurement prevents false measurements
− Automatically applies three testing frequencies for the most
accurate readings
− Detects faulty test conditions such as poor soil conditions and
input noise
PHASE SEQUENCE
− Phase sequence indication
− Frequency measurement
− Phase-to-Phase voltage measurement
Amprobe® Test Tools
DM-III MULTITEST F Power Quality Recorder
Specifications
Supplied Current Transducer
Input accuracy
AC Current
AC Voltage including
Sags and Surges
Harmonics
Power
Energy
Peak Demand
Power Factor
Frequency measurement
Phase sequence
Co-generation
Selectable Fundamental
Frequencies
Available Recording Time
Megohmmeter
Insulation resistance with
50 VDC test voltage
Insulation resistance with
100 VDC test voltage
Insulation resistance with
250 VDC test voltage
Insulation resistance with
500 VDC test voltage
Insulation resistance with
1000 VDC test voltage
Data Sheet
AM-FLEX33 3000A Flexible 7” internal diameter CT
±(0.5% Rdg + 2 LSD)
DM-CT-100: 0.5A to 100A; DM-CT-HTA: 5 – 1000A; AM-FLEX33: Selectable: 5 – 1000A or 15 – 3000A
±(0.5% Rdg + 2 LSD)
±(0.5% Rdg + 2 LSD) @DC to 25 harmonics**
Working (W), Reactive (VAR) and Apparent (VA) ±(1.0% Rdg + 2 LSD)
Working (kWh), Reactive (VARh) and Apparent (VAh) ±(1.0% Rdg + 2 LSD)
KW ±(1.0% Rdg + 2 LSD)
0.00 – 1.00
57 to 63.6 Hz at 60Hz fundamental; 47 to 53 Hz at 50Hz fundamental; ±(1.0% Rdg + 2 LSD)
1-2-3
Computes incoming and outgoing energy
50/60 Hz
Several hours to several years depending on setup
Range
0.01 – 19.99, 49.9
49.9 – 99.9MΩ
0.01 – 19.99, 99.9
99.9 – 199.9MΩ
0.01 – 19.99, 199.9, 249
249 – 499 MΩ
0.01 – 19.99, 199.9, 499
499 – 999 MΩ
Accuracy
± (2% Reading + 2 digits)
± (5% Reading + 2 digits)
± (2% Reading + 2 digits)
± (5% Reading + 2 digits)
± (2% Reading + 2 digits)
± (5% Reading + 2 digits)
± (2% Reading + 2 digits)
± (5% Reading + 2 digits)
± (2% Reading + 2 digits)
± (5% Reading + 2 digits)
± (2% Reading + 2 digits)
± (2% Reading + 2 digits)
Accuracy
± (5% Reading + 3 digits)
± (5% Reading + 3 digits)
± (5% Reading + 3 digits)
0.01 – 19.99, 199.9, 999
999 – 1999 MΩ
Low Resistance (without timer)
0.01 – 19.99, 99.9Ω
Low Resistance (with timer)
0.01 – 9.99Ω
Ground Resistance
Ground resistance
Ground resistivity
Range
0 – 19.99, 199.9, 1999
Ω
0.6 – 125.6
Ωm
0.125 – 1.256, 19.99, 199.9 kΩm
LowΩ: 200mA Continuity Test (AUTO, RT+, RT- Mode)
Range [Ω]
Resolution [Ω]
Accuracy(*)
0.01 – 9.99
0.01
±(2% Reading + 2 digit)
10.0 – 99.9
0.1
±(2% Reading + 2 digit)
(*) After Test leads calibration
Test Current
> 200mA DC per R≤5Ω (Test leads included)
Resolution for Test current
1mA
Open Circuit Voltage
4V
V0
24V
www.Amprobe.com
DM-III MULTITEST F Power Quality Recorder
Specifications
(continued)
Insulation Test
Test Voltage [V]
50
Range [MΩ]
0.01 – 9.99
10.0 – 49.9
50.0 – 99.9
0.01 – 9.99
10.0 – 99.9
100.0 – 199.9
0.01 – 9.99
10.0 – 199.9
200 – 249
250 – 499
0.01 – 9.99
10.0 – 199.9
200 – 499
500 – 999
Resolution [MΩ]
0.01
0.1
0.1
0.01
0.1
0.1
0.01
0.1
1
1
0.01
0.1
1
1
Accuracy
±(2% Reading + 2 digit)
±(2% Reading + 2 digit)
±(5% Reading + 2 digit)
±(2% Reading + 2 digit)
±(2% Reading + 2 digit)
±(5% Reading + 2 digit)
±(2% Reading + 2 digit)
±(2% Reading + 2 digit)
±(2% Reading + 2 digit)
±(5% Reading + 2 digit)
±(2% Reading + 2 digit)
±(2% Reading + 2 digit)
±(2% Reading + 2 digit)
±(5% Reading + 2 digit)
±(2% Reading + 2 digit)
±(2% Reading + 2 digit)
±(2% Reading + 2 digit)
±(5% Reading + 2 digit)
Data Sheet
100
250
500
1000
Open circuit Test Voltage
Short Circuit Current
Nominal Test Current
0.01 – 9.99
0.01
10.0 – 199.9
0.1
200 – 999
1
1000 – 1999
1
<1.3 x Nominal Test Voltage
<6.0mA with 500V Test Voltage
500V: >2.2mA with 230kΩ;
other: >1mA with 1kΩ*Vnom
Frequency Measurement
Range [Hz]
Resolution [Hz]
Accuracy
47.0 – 63.6
0.1
±(0.1%Reading+1 digit)
RCD and LOOP function are active only for 50Hz ± 0.5Hz frequency
Phase Rotation: Voltage Measurement
Range [V]
Resolution [V]
0 – 460V
1
Ground Test: Resistance Measurement With Earth Rods
Range RE [Ω]
Resolution [Ω]
0.01 – 19.99
0.01
20.0 – 199.9
0.1
200 - 1999
1
Accuracy
±(5% Reading + 3 digit)
Test Current
<10mA – 77.5Hz
Open circuit Test Voltage
<20V RM
Accuracy
±(3%Reading + 2 digit)
Amprobe® Test Tools
DM-III MULTITEST F Power Quality Recorder
Specifications
(continued)
Ground Test: Resistivity Measurement
Range p
Resolution
0.60 – 19.99
Ωm
0.01
Ωm
20.0 – 199.9Ωm
0.1Ωm
200 – 1999Ωm
1
Ωm
2.00 – 99.99kΩm
0.01 kΩm
100.0 – 125.6kΩm(*)
0.1 kΩm
(*) setting distance = 10m
Accuracy
±(5% Reading + 3 digit)
Test Current
<10mA – 77.5Hz
Open circuit Test Voltage
<20V RMS
Voltage Measurement – (Autorange)
Range [V]
Resolution [V]
15 – 310V
0.2V
310 – 600V
0.4V
Accuracy
±(0.5% Reading+2digit)
Voltage Sag And Surge Detection – (Manual Range)
Range [V]
Resolution (Voltage)
15 – 310V
0.2V
30 – 600V
0.4V
Resolution (Time)
10ms (_ period)
Accuracy (Voltage)
±(1.0% Reading+2digit)
Accuracy (Rif. 50hz) (Time)
10ms (_ period)
Input Impedance
300kΩ (Phase-Neutral); 300kΩ (Phase-Phase)
Current Measurement – STD & FlexEXTclamps
Range [V]
Resolution [Mv]
0.005 – 0.26V
0.1
0.26 – 1V
0.4
(*): Example: with a 1000A/1V full scale clamp, the instrument detect only current higher than 5A
Accuracy
±(0.5% Reading+2digit)
Input Impedance
200kΩ
Overload Protection
5V
Current Measurement – FlexINT clamp – 1000A Range
Current Range
Input Voltage Range
5.00 – 20.00A
425µV – 1.7mV
20.00 – 99.99A
1.7mV – 8.499mV
100.0 – 999.9A
8.5mV – 84.99mV
Input Impedance
9.166kΩ
Overload Protection
5V
Data Sheet
Resolution
0.850µV
0.850µV
8.5µV
Accuracy
± (4.0%rdg + 8.5µV)
± (1.0% rdg + 8.5µV)
± (1.0% rdg + 85µV)
www.Amprobe.com
查看更多>
参数对比
与DM-III MULTITEST相近的元器件有:DM-III MULTITEST F。描述及对比如下:
型号 DM-III MULTITEST DM-III MULTITEST F
描述 power analyzers pwr qual record/anal data logger power analyzers pwr qual record/anal data logger flex
下载资源,评论返积分啦!
网友们以后对下载过的资源进行评价,可以返积分啦! 评价一个返一分,一天最高能返5分。 希望网友们积极...
hightemplar 下载中心专版
网桥的功能
1.源地址跟踪   网桥具有一定的路径选择功能,它在任何时候收到一个帧以后,都要确定其正确的传输路径...
mdreamj RF/无线
求合作微波射频工程师
我们有个机顶盒样的东西,现在需要抄板子,把PCB电路图,电子元器件详细清单还有工艺书弄出来,有那位感...
yineng09 RF/无线
楼宇对讲音频失真怎么办
在楼宇对讲工程中,出现了视频会影响音频的质量的情况,请问此种情形该如何处理? 楼宇对讲音频失真怎么办...
xyh_521 安防电子
世界各国电视制式
PDF格式的哦 世界各国电视制式 本帖最后由 jameswangsynnex 于 2015-3...
feifei 移动便携
反激电源几个重要的公式的总结
反激电源设计中有几个比较重要的公式,下面给大家列一列: ...
okhxyyo 电源技术
热门器件
热门资源推荐
器件捷径:
S0 S1 S2 S3 S4 S5 S6 S7 S8 S9 SA SB SC SD SE SF SG SH SI SJ SK SL SM SN SO SP SQ SR SS ST SU SV SW SX SY SZ T0 T1 T2 T3 T4 T5 T6 T7 T8 T9 TA TB TC TD TE TF TG TH TI TJ TK TL TM TN TO TP TQ TR TS TT TU TV TW TX TY TZ U0 U1 U2 U3 U4 U6 U7 U8 UA UB UC UD UE UF UG UH UI UJ UK UL UM UN UP UQ UR US UT UU UV UW UX UZ V0 V1 V2 V3 V4 V5 V6 V7 V8 V9 VA VB VC VD VE VF VG VH VI VJ VK VL VM VN VO VP VQ VR VS VT VU VV VW VX VY VZ W0 W1 W2 W3 W4 W5 W6 W7 W8 W9 WA WB WC WD WE WF WG WH WI WJ WK WL WM WN WO WP WR WS WT WU WV WW WY X0 X1 X2 X3 X4 X5 X7 X8 X9 XA XB XC XD XE XF XG XH XK XL XM XN XO XP XQ XR XS XT XU XV XW XX XY XZ Y0 Y1 Y2 Y4 Y5 Y6 Y9 YA YB YC YD YE YF YG YH YK YL YM YN YP YQ YR YS YT YX Z0 Z1 Z2 Z3 Z4 Z5 Z6 Z8 ZA ZB ZC ZD ZE ZF ZG ZH ZJ ZL ZM ZN ZP ZR ZS ZT ZU ZV ZW ZX ZY
需要登录后才可以下载。
登录取消